The
list gives an overview of the compatibility of the RS 232 interface at all
h/p/cosmos running machines to ECG-, Ergospirometry systems, Blood-pressure monitors and PC software programs.
Beside this a directly
connected printer is possible as well. Analog communication is
possible with the interface converter, which is optionally available. The
protocol h/p/cosmos
coscom for the RS232 interface and the coscom.dll are available for
download. |
Always
use IEC 601-1 approved potential isolation components when linking medical
devices via interface.
Also available for adaptation to PC, ECG or
Ergospirometry systems
is a small and handy running machine-simulator with key board and
display, which simulates all functions of the h/p/cosmos running machines
controlled via RS232 interface. |
In
case of any further questions or for service enquiry, please fill out the RS232
connection form |
|
h/p/cosmos only approves the interface protocol „coscom
v3“ for medical applications and herewith confirms the -since
March 21, 2010 (MDD / Directive 2007/47/EC) mandatory- conformity of
coscom v3 according to the norms EN 62304 (Life Cycle Requirements for
Medical Device Software; Medical Device Software Life Cycle Processes)
and EN 14971 (risk management for medical devices) as well as the
compliance to a very high safety standard.
All other interface protocols (older versions coscom v2,
coscom v1, coscomekg.dll, Trackmaster, Pacer, etc.) are not validated by
h/p/cosmos according to the norms EN 62304 and EN 14971 and must not be
used for medical applications (e.g. ergometry, stress-tests,
cardiorespiratory diagnostic, performance diagnostic, etc.) since March
21, 2010 in connection with h/p/cosmos treadmills or h/p/cosmos OEM
treadmills!
We strongly recommend to update to the latest h/p/cosmos
coscom v3 interface protocol. The latest coscom.dll and implementation
notes can be found on www.coscom.org. In case you have any questions please contact
service@h-p-cosmos.com |