Interface converter - RS232 / analog and USB / RS232 are available as an optional equipment.
Elevation and speed can be
controlled by an external ECG or Ergospirometry with analogue voltages of
0 - 5 Volts (or 0 – 10 Volts) at the separate socket.
Furthermore
analogue output signal for elevation and speed of the running machine as
well as an analogue signal for the heart rate of the subject is available
for host equipment. So with this interface converter a complete analogue
communication is possible.
[ cos10056 interface converter cable for
printer
RS232 - Centronix seriell - parallel
]
Beside this a directly
connected printer is possible as well. Analog communication is
possible with the interface converter, which is optionally available. The
protocol h/p/cosmos
coscom for the RS232 interface and the coscom.dll are available for
download at www.coscom.org |
 [
cos12769-01 interface converter
adapter USB / RS232]
[new version please download driver
here!]

[
cos12769 interface converter cable USB / RS232]
[version year 2000 - 2007 please download driver
here!]
A
list gives an overview of the compatibility of the RS 232 interface at all
h/p/cosmos running machines to ECG-, Ergospirometry systems, Blood-pressure monitors and PC software programs.

For any interface connections for medical application a potential
isolation based on IEC 60601-1 must be established between the interface
ports of the running machine and the host device. For example via
opto-coupler.
Always
use IEC 601-1 approved potential isolation components when linking medical
devices via interface.
|

[
cos10223 Potential equalization cable
]
Also available for adaptation to PC, ECG or
Ergospirometry systems
is a small and handy running machine-simulator with key board and
display, which simulates all functions of the h/p/cosmos running machines
controlled via RS232 interface.
In
case of any further questions or for service enquiry, fill out the RS232
connection form |